Probes on a calibration substrate (FormFactor)
TECH GUIDE / DSF SYSTEM

Find answers to your questions
about probes and calibration.

Explore frequently asked questions for product guidance and related technical resources.

Videos
50
FAQ
47
MOST ASKED / TOP 5

Most asked

Start with the questions we hear most often. Select one to jump to its answer.

01 / PROBE SELECTION

Probe selection & pad design

8 questions
QWhat should I check when choosing an RF probe?

Look at the maximum frequency, connector, tip pitch, contact configuration (GSG, GS, SG and so on), pad metal and measurement environment together. FormFactor covers the electrical selection criteria, from connector type to tip pitch and architecture, in a blog post and the Probe Selection Guide.

  1. Confirm the highest frequency you need and the connectors on your VNA or extenders.
  2. Check the DUT pad configuration (GSG, GS, SG, GSSG) and pitch in µm.
  3. Decide the pad metal (Al or Au) and the environment, such as temperature or vacuum.
  4. Use the Probe Selection Guide to pick the matching probe family.
QHow do Infinity, ACP and |Z| probes differ?

The three families differ in construction and strengths. Choose by your pads and environment.

  1. Infinity: lithographic thin-film construction with low contact resistance even on aluminum pads (typically < 0.05 Ω on Al, < 0.02 Ω on Au) and low crosstalk, suited to device characterization and modeling. Connectors 40–145 GHz, pitch 50–250 µm.
  2. ACP: a rugged general-purpose probe with a compliant Air Coplanar tip that handles non-planar surfaces, with a typical life of about 500,000 contacts on gold pads. DC–110 GHz, -65 to +200 °C, pitch 50–1250 µm.
  3. |Z|: the signal makes a single transition inside a shielded, air-isolated body, suiting 4 K to 300 °C environments; independent contacts handle pad height differences up to 50 µm and need less overtravel.

Figures are from FormFactor product pages.

QHow do InfinityXT and InfinityXF differ from Infinity?

InfinityXT extends the Infinity family, while InfinityXF is a newer probe for coaxial broadband measurement from DC up to 250 GHz. Compare frequency, pitch and lifetime details in each data sheet.

QWhich probe suits small pads or multi-contact measurement?

FormFactor offers InfinityQuad, a configurable multi-contact RF/mmW probe for small pads, citing pads as small as 30 × 50 µm. The configuration depends on your signal layout and is specified with a design capture form.

QWhere can I get probe S-parameter (S2P) data?

From the “ACP/Infinity S2P Download” and “|Z| Probe Device Data” sections of FormFactor’s Analytical Probe Support page.

QHow do I order a custom probe?

FormFactor provides specification forms for each probe type. Fill in the relevant form and send it with your technical inquiry.

  1. Download the custom probe order form or the design capture form for the probe family.
  2. Fill in the pad layout, pitch, frequency and other requirements.
  3. Attach the completed form to a technical inquiry.
QWhere are the pad layout rules for GHz probing?

FormFactor publishes “Layout Rules for GHz-Probing” and “Infinity Probe Mechanical Layout Rules”. Two blog posts also explain the layout considerations for Infinity probes.

QWhat are the pad layout rules for InfinityQuad?

InfinityQuad follows its own document, “InfinityQuad ProbePAD Layout Rules”.

02 / STATION SETUP

Station setup & installation

8 questions
QWhat is the setup sequence for two-port RF measurement on a manual station?

FormFactor’s nine-part RF Probe Video Guide follows this sequence, with a short video for each step.

  1. Arrange the station for two-port microwave measurement.
  2. With the platen handle fully down, mount the probes on the positioners and keep it down until setup is complete.
  3. Place the contact substrate (gold on alumina) for planarity checks and the ISS on the chuck.
  4. Focus the microscope on the contact substrate and lower the tips with the Z micrometer to planarize.
  5. With the probes clear of the substrate, move the stage over the ISS.
  6. Align to the left ISS marker, move only in X, and check the probes sit the same way on the right marker.
  7. Use the back of the ISS map to locate the short, load, thru and line structures.
  8. Make first contact at reference plane X and set the skate.
  9. Raise and lower the platen while moving in X to contact load, short and thru in turn; measure the open with the probes raised.

The videos were filmed on an older Summit 9100 manual station.

QHow do I set skate (overtravel)?

Skate is the forward movement of the tips after first contact. Part 8 of the RF Probe Video Guide shows making first contact at reference plane X (midway between the outer flat edge and the inner apex) and then setting skate. Recommended values vary by probe, so check the probe’s documentation.

QHow do I check the probes are aligned to the ISS?

Use the alignment markers on both sides of the ISS.

  1. Place the probes over the marker in the left corner of the ISS.
  2. Move the stage in X only to the right-hand marker.
  3. If the probes sit the same way on the right marker they are aligned; if not, realign them.
QHow do I choose positioners?

FormFactor offers manual and motorized positioners for DC, RF and optical measurement; compare them in the Probe Positioners Brochure.

QHow do I choose between 150, 200 and 300 mm manual stations?

Start with wafer or substrate size: FormFactor’s manual stations are the MPS150 for 150 mm; PM8, Summit, EPS200RF and EPS200MMW for 200 mm; and PM300 for 300 mm. Then configure for RF, mmW or thermal measurement.

QWhich microscope (eVue) is used on the stations?

FormFactor’s digital microscopes for its stations are the eVue family; the current data sheet is for eVue V.

QWhere do I find facility requirements before installation?

Each model’s Facility Planning Guide lists what to prepare in your facility before the system arrives.

QHow do I prepare for vibration isolation and EMI shielding?

Installation requirements are in the vibration isolation and shield enclosure documents.

03 / ISS · CAL SUBSTRATE

ISS & calibration substrates

5 questions
QWhere do I find ISS calibration coefficients (Co, Ls, Lterm)?

The coefficients are supplied with the probe, not the ISS, because the electrical behavior of the standards depends on the probe and its placement.

  1. Find the coefficients on the data sheet shipped with the probe.
  2. Or open Tools → View Probes and Matching ISS/CSR in WinCal.
  3. Enter the three values per probe (open capacitance Co, short inductance Ls, load inductance Lterm) and the thru delay into the cal kit.
QWhat thru delay should I enter?

1 ps on a standard-pitch ISS with probe spacing set by the alignment marks, and 4 ps on a wide-pitch ISS.

250 µm pitch probes can use either ISS, but the supplied coefficients refer to the standard-pitch ISS.

QHow do I find the electrical length (delay) of ISS lines?

TRL needs the delay of the long lines entered in WinCal or the VNA. Derive it from the drawing supplied with the ISS.

  1. Measure the line length on the ISS drawing using the scale provided.
  2. Allow for about 50 µm of probe overlap at each end.
  3. Convert at about 130 µm ≈ 1 ps: delay (ps) = length (mm) ÷ 0.13.

Example: the 900 µm line on the 101-190 ISS measured 6.6 ps after an LRRM calibration.

QWhich ISS matches my probe?

Choose an ISS by tip configuration (GSG, GS/SG, GSGSG and so on), pitch range and recommended upper frequency. WinCal’s Tools → View Probes and Matching ISS/CSR also shows the matching ISS or CSR for your probe.

  1. Check your probe’s tip configuration and pitch.
  2. Find substrates for that configuration and pitch range in FormFactor’s ISS table.
  3. Pick by upper frequency, for example GSG 100–250 µm: 101-190 (67 GHz) or 143-033 (145 GHz); GSG 50–75 µm: 138-356 (325 GHz). 162-641 is a verification-only substrate.

|Z| probes use CSR substrates and T-Wave probes use multiline TRL substrates.

QHow do ISS, CSR and multiline TRL substrates differ?

They are designed for different probe families and differ in construction.

  1. ISS: for ACP, Infinity and FPC probes, built around LRRM calibration with automatic load inductance compensation.
  2. CSR: for |Z| probes, with hard-gold standards for long life, an on-substrate open standard, and loads trimmed better than 0.1 % that stay within 0.3 % from 10 to 430 K.
  3. Multiline TRL: mm- and sub-mm-wave substrates for T-Wave probes with CPW short, thru and two lines, recommended for use with WinCal (172-885/886/887).
04 / CALIBRATION METHODS

Calibration methods & troubleshooting

12 questions
QWhy do SOLT and LRRM give different coefficients?

This is normal: the two methods assume different things about the standards.

  1. SOLT assumes the short, open, load and thru behavior is fully known and enters it as a cal-kit model.
  2. On-wafer, probe placement on the standards varies (operator positioning is typically no better than 5 µm), so real behavior departs from the model and the difference goes into the error terms.
  3. LRRM needs only a fully known line (thru) and a match standard with known DC resistance and constant inductance; the reflects just need to be equal on both ports.
  4. Even with the same standards, LRRM assumes less, so WinCal (LRRM) and a VNA’s built-in SOLT give different coefficients.

Both methods assume a single propagation mode, so neither is perfect.

QWhen is SOLR used, and how do I enter the thru delay?

SOLR is used when a well-known thru is not possible, such as double-sided device probing. The thru delay only selects the sign of a root, so a rough estimate is enough.

  1. Enter an estimated delay for the thru (R) standard; FormFactor’s FAQ lists estimates for the lines on the 005-016 ISS.
  2. Run the calibration.
  3. Look at the S21 phase of the thru standard on a polar plot.
  4. If the phase is continuous with frequency, the result is good; if it steps, adjust the estimate and recalculate.
QHow do I get an accurate TRL or multiline TRL calibration?

TRL needs the delay (electrical length) of the line standards in WinCal or the VNA, and custom lines need an accurately known characteristic impedance Z0.

  1. Derive the line delays from the ISS drawing (about 130 µm ≈ 1 ps).
  2. For custom lines, measure and set the characteristic impedance Z0.
  3. For multiline TRL, use the dedicated substrates (172-885 to 887) and their maps.
QHow is a 4-port on-wafer calibration done?

FormFactor uses a hybrid 4-port calibration that combines LRRM and SOLR.

  1. In 4-port setups some thrus become loop-back or crossover paths because of probe geometry and can differ greatly from the classic thru definition.
  2. More ports mean more thru combinations to measure and more risk of alignment error.
  3. LRRM does not need known shorts or opens (only symmetry), so it is less sensitive to scrub variation, while SOLR accepts a reciprocal thru; the hybrid uses both strengths.
QHow is crosstalk in dual-signal (GSSG) probes corrected?
QHow are multi-channel probes calibrated?

A FormFactor paper proposes adding one GSG probe to measure thrus for multi-channel probes placed in several quadrants. This allows a conventional calibration without a custom cal kit, and the delay inconsistencies are shown to have minimal effect on uncertainty in most practical cases.

QWhy use probe-tip power calibration?

FormFactor papers report that power calibration referenced to the probe tip improves the accuracy and stability of wafer-level S-parameter measurements (to 110 GHz, for 5G applications).

QHow can I reduce variation in fine-pitch RF calibration?

FormFactor covers how fine-pitch RF calibration affects wafer test accuracy above 80 GHz in a paper and a blog post.

QWhy does “Calibration no longer acceptable” appear right after calibrating?

It indicates poor repeatability in the measurement system.

  1. Check the cal-kit definition (for example, that a short was not measured in place of an open).
  2. With calibration off, compare repeated raw measurements to check system repeatability.
  3. Follow FormFactor’s FAQ for the remaining checks.
QWhy is there a 180° step in S21 phase after SOLR?

The SOLR algorithm chose the wrong root. Adjust the thru delay estimate and recalculate to remove the step.

  1. Adjust the thru delay estimate.
  2. Recalculate the calibration.
  3. Check that the thru standard’s S21 phase is continuous.
QHow do I validate a calibration?

WinCal includes validation tools. See “Calibration Validation and System Comparison Tools.pdf” under Help → Documentation Folders → Application Toolkits.

QHow do I keep calibration accurate at high and low temperatures?

A FormFactor workshop covers setup, calibration and performance evaluation for coaxial (to 145 GHz) and waveguide (WR15 to WR1) probes from -40 to 125 °C.

05 / WINCAL

Using WinCal

9 questions
QHow do I activate or deactivate a WinCal license?

In WinCal 4.9 and later, launch WinCal and use Help → Licensing.

Online activation errors in 4.7 and earlier were resolved by upgrading.

QHow do I connect WinCal to a VNA?

FormFactor’s “WinCal Connecting to a VNA with VISA Options” describes the connection settings.

Over GPIB, station control software (Nucleus) sharing the bus can conflict; use a GPIB switch box or disable GPIB control in Nucleus.

QCan I turn off the prompt for every measurement on a manual prober?

Yes. Use the checkbox on the Calibration tab in Options.

  1. Open Options in WinCal.
  2. Go to the Calibration tab.
  3. Set the checkbox for the per-measurement prompt.

Menu names follow FormFactor’s FAQ and may differ slightly in WinCal 5.

QCan I export 12-term error coefficient files?

Yes. Advanced calibrations such as LRRM, LRM+ and TRL are stored as 16-term sets (in the ErrSets folder) and can be converted on command to 12-term and saved as one S1P file per coefficient.

QWhere do I see calibration data and reports?

The View button on the calibration form opens the measured data in a Report. Example reports can be copied with Help → Copy Examples to My Documents.

QWhat changed in WinCal 5 and 5.1?
QCan I create a custom wizard?

Yes. Use Wizards → Wizard Script Editor, or copy an existing wizard (Create Copy for User / Group) and modify it.

QHow do I keep the calibration setup and raw data?

Save the .WCF file; WinCal does not save it by default.

  1. After calibrating, save the .WCF file.
  2. It stores the calibration setup and the raw data used to create the calibration.
  3. You can later un-correct data, change the algorithm and re-correct, which helps with fault-finding and traceability.
QWhich calibration methods does WinCal support?

According to FormFactor, WinCal supports 1-, 2-, 3- and 4-port algorithms including eLRRM, LRM+, SOLR-LRRM hybrid and NIST-style multiline TRL, plus Error Set Management for comparing and augmenting error sets.

This describes WinCal XE; check the WinCal data sheet for current features.

06 / PROBE CARE

Probe care & repair

5 questions
QHow do I clean probe tips?

FormFactor shows three methods in separate videos: cleaning pad, compressed air and cleaning brush. Infinity probes have their own recommended procedure.

QHow do I planarize probes?

Adjust the probe so all tips touch the contact substrate together.

  1. Place the contact substrate (gold on alumina) on the chuck.
  2. Focus the microscope on the substrate surface.
  3. Slowly lower the tips with the Z micrometer and watch them come into focus.
  4. Follow the video for checking and adjusting planarity.
QHow do I mount and remove probes on frequency extenders?
QWhat are the repair limits and eligibility conditions for probes?

Based on FormFactor’s Analytical Probe Repair Program.

  1. Listed, currently manufactured probes are repaired at a fixed price.
  2. ACP, Infinity and similar probes can be repaired up to three times; serial numbers ending in “-3” are not eligible.
  3. |Z| probes are not covered by this program.
  4. T-Wave probes with undamaged blocks can be repaired without limit.
  5. Obsolete products are excluded, and probes older than five years are evaluated first.

Prices and lead times follow FormFactor’s page. For submitting through DSF, please use Repair Request.

QHow do I store probes when changing setups?

FormFactor RF arm storage pods are used to store probes when changing measurement setups.

07 / mmW · THz · DC

mmW, THz & DC measurement

5 questions
QHow do I measure several mmW bands on one station?

The EPS200MMW uses SIGMA integration kits to switch band setups on a single manual station.

QInclined or horizontal waveguide port savers: which is better?

FormFactor makes extensive use of inclined extenders for both coaxial single-sweep and banded waveguide setups.

  1. Inclined extenders suit thermally optimized platens with insulation layers, giving good positioning stability.
  2. They sit farther from the hot chuck, which helps temperature stability.
  3. Comparisons with direct horizontal connection had been limited to 220 GHz at ambient; the webinar compares both in the WR3.4 band at ambient and over temperature with T-Wave and waveguide Infinity probes.

Which is better depends on conditions; see the webinar for the results.

QHow do I measure broadband to 220 GHz in a single sweep?
QWhat are the main challenges of sub-THz measurement?
QHow can I improve DC accuracy for device modeling?
NEXT / SUPPORT

Didn’t find it?
Ask our engineers.